AUTOMATIC DISTILLATION OF PETROLEUM PRODUCTS AT ATMOSPHERIC PRESSURE

NDI 450 Normes : ASTM D 1078, ASTM D 850, ASTM D 86, IP 123, ISO 3405

DESCRIPTION

Reliable and robust, the NDI 450 is a fully automated distillation unit at atmospheric pressure with a combination of latest technologies. Due to our expertise, the NDI 450 offers you the best performance thanks to its low voltage heating element (easily removable) and its built-in Peltier cooling system.
Developed with a windows software preprogrammed with standard methods, the NDI 450 offers the possibility to create your own methods (more than 1000 methods can be stored). All parameters can be followed during the test.





TECHNICAL DATA
Power Supply 230 V, 50/60 Hz – 9A
Dimensions 54cm/W 45cm/D 63cm/H
Weight 70 kg

CHARACTERISTICS


  • Temperature measurement range : Ambient to 450°C
  • Temperature accuracy : 0.05°C
  • Condenser temperature : 0 to 60°C
  • Built-in Cooling System
  • Level follower accuracy : 0.1 ml with 0,05 ml resolution
  • Receiver compartment temperature : 1 to 60°C
  • Quick cooling of heater by fan
  • Low voltage heating element offering 850W power and controlled by thermocouple


RECOMMENDATIONS
Nitrogen technical quality (for filling the extinguishing system)
AN UPS IS RECOMMENDED
Nitrogen technical quality (for filling the extinguishing system)
AN UPS IS RECOMMENDED

FEATURES


  • For a complete range of analysis, the NDI 450 is ready to use as per ASTM D86 and related methods (Optional kit for ASTM D850, D1078), also including an automatic cetane index calculation (ISO 4264). In addition to unique performance, the NDI 450 owns all safety features which make it the most secure distillation.
  • In addition to an automatic and manual N2 fire extinguisher (N2 empty bottle delivered with the unit), the NDI 450 is equipped of 6 sensors that can automatically stop the test in case of danger :
    • Receiver position
    • Compartment door closed
    • Drip deflector in position
    • PT 100 probe in position
    • Coolant liquid : low level detection
    • N2 minimum pressure required